Cleanliness testing laboratory
keyword:
Testing laboratories
Classification:
Laboratory
Product Features:
National consultation hotline:
- DESCRIPTION
- Technical Parameter
- Product Details
-
Harco Technology provides users with complete solutions for cleanliness analysis laboratories, including complete process services such as program consultation, equipment configuration, site planning, installation and commissioning, personnel training, and after-sales maintenance.
Laboratory environment requirements
Laboratory site area
It is recommended to be about 20㎡, the ground is flat and dust-free, and the air outlet of Φ100mm is reserved
Lab type
Class 100,000 clean room or separate partition room
Lab environment
No dust particles, vibration sources
Laboratory electricity requirements
Power: AC220V±10%, frequency: 50Hz, power: 4Kw, socket needs to be grounded
0.2-0.4Mpa pressure drying air source
Lab door size
The height is greater than 2m and the width is greater than 1m
Other laboratory requirements
3m*0.65m*0.8m experimental table, balance shockproof table